BSC: Bio-Safety Cabinet

Description

  • This is a completely closed cabinet, isolate operating personnel or objects to avoid damage by harmful substances, protective role.
  • It is often used in high-polluting, high toxicity, high hazard job.
  • Exposure to toxic chemical biology, food testing, pharmaceutical and health hazardous industries ideal for devices.
  • Customized size and specification are available to meet customer's requirements.
  • EN 12469 Certified (optional)

The categories of Bio Safety Cabinet are defined and classified according to the National Sanitation Foundation Standard, NSF/ANSI 49.

Class I

The Class I BSC is designed to provide personnel protection

BSC grades

Inlet air velocity

% of air flow

Exhaust system

Applications

m/s

internal
re-circulated

Exhausted

Class I

0.38

0

100

External
(hard duct)

It is used for laboratories that work with
nonvolatile, radionuclide/chemical
protection, and microorganisms at Risk.


Class II: A1, A2, B1, B2

The Class II BSC is designed not only to provide personnel protection but also to protect work surface materials from being contaminated by the room air which is drawn into the cabinet. And it is used for protecting the laboratory environment as well. It is acceptable for working with medium-level risk tests. According to the National Sanitation Foundation Standard, NSF/ANSI 49(2002), it can be defined and classified into 4 types. Their features are compared below.

BSC grades

Inlet air velocity

% of air flow

Exhaust system

Applications

m/s

internal
re-circulated

Exhausted

Class II
A1

0.38

70

30

Internal or

external

(Exhaust hood)

It is used for laboratories which are work
with nonvolatile, radionuclide/chemical
protection,microorganisms in Risk.

Class II
A2

0.5

70

30

Class II
B1

0.5

30

70

External
(Hard duct)

It is used for laboratories which are work
with lightly volatile, radionuclide/chemical
protection,microorganisms in Risk.

Class II
B2

0.5

0

100


Class III

Class III biological safety cabinets are suitable for work in Bio-safety Level 3 and 4 laboratories and provide protection for the highest hazardous level of Risk Group 3, Risk Group 4, or above. It virtually is a sealed gas-tight cabinet. Moreover, its supply air is HEPA- filtered, and exhaust air passes through two HEPA filters, which keep the cabinet interior under negative pressure consequently. The only way of accessing the work surface for operation is by means of heavy-duty rubber gloves, which are attached to ports in the cabinet virtually.

BSC grades

Inlet air velocity

% of air flow

Exhaust system

Applications

m/s

internal
re-circulated

Exhausted

Class III

N/A

0

100

Exhaust air should pass through two filters.

It is used for laboratories that work with
lightly volatile, toxic, radionuclide-protection
microorganisms at Risk.


Selection of a biological safety cabinet (BSC)

BSC category

Laboratory type

Type of protection

Class I

Groups 1–3, microorganisms in Risk;
Slightly volatile, radionuclide/chemical protection

Operator, environment

Class Ⅱ A1

Groups 1–3, microorganisms in Risk

Operator, environment, product

Class Ⅱ A2

Groups 1–3, microorganisms in Risk

Operator, environment, product

Class Ⅱ B1

Groups 1–3, microorganisms in Risk;
Slightly volatile radionuclide/chemical protection

Operator, environment, product

Class Ⅱ B2

Groups 1–3, microorganisms in Risk;
Minute amounts volatile, radionuclide/chemical protection

Operator, environment, product

Class III 

Groups 1–3, microorganisms in Risk;
Volatile, radionuclide/chemical protection

Operator, environment, product

 

Type-CLASS IIA1A2B1B2A1A2B1B2A1A2B1B2
Model NumberBA1-4BA2-4BB1-4BB2-4BA1-5BA2-5BB1-5BB2-5BA1-6BA2-6BB1-6BB2-6
External Dimension
( W * D * H mm)
1450*827*22001750*827*22002050*827*2200
Internal Dimension
( W * D * H mm)
1250*600*6501550*600*6501850*600*650

Product meets
Standard

NSF 49、EN12469、JIS K 3800

Air cleanliness
standard
(Within working
area)

ISO 14644-1
Body Construction1.5 t SUS 304
HEPA filter
0.3 um H14
1170*450*70 mm 1470*450*70 mm 1770*450*70 mm 
HEPA filter
(Exhaust)
520*460*100 mm
0.3 um H14
520*460*290 mm
0.3 um H14
700*460*100 mm
0.3 um H14
700*460*292 mm
0.3 um H14
1000*460*100mm
0.3 um H14
1000*460*292 mm
0.3 um H14
BlowerEC Brushless Motor
WorktableSUS 304 perforated panelSUS 304 perforated panelSUS 304 perforated panel
Air in flow75 FPM100 FPM100 FPM100 FPM75 FPM100 FPM100 FPM100 FPM75 FPM100 FPM100 FPM100 FPM
Air down flow60 FPM60 FPM60 FPM
Air circulation70%70%30%0%70%70%30%0%70%70%30%0%
Air Exhaust30%30%70%100%30%30%70%100%30%30%70%100%
Light Illuminance
( NSF 49 )
> 600 lux> 600 lux> 600 lux

Noise Level
(According to
EN12469)

< 65 dBA< 65 dBA< 65 dBA
UV Lamp254nm254nm254nm

Biosafety cabinet (BSC): class II A2
Dimensional structure drawing for Biosafety cabinet (BSC): class II type B2